Sec. 15.241 Operation in the band 174–216 MHz.
(a) Operation under the provisions of this section is restricted to
biomedical telemetry devices.
(b) Emissions from the device shall be confined within a 200 kHz band which
shall lie wholly within the frequency range of 174–216 MHz.
(c) The field strength of any emissions radiated within the specified 200
kHz band shall not exceed 1500 microvolts/meter at 3 meters. The field
strength of emissions radiated on any frequency outside of the specified 200
kHz band shall not exceed 150 microvolts/meter at 3 meters. The emission
limits in this paragraph are based on measurement instrumentation employing
an average detector. The provisions in Sec. 15.35 for limiting peak emissions
apply.