Goto Section: 15.240 | 15.242 | Table of Contents

FCC 15.241
Revised as of October 1, 2007
Goto Year:2006 | 2008
Sec.  15.241   Operation in the band 174–216 MHz.

   (a)  Operation  under  the provisions of this section is restricted to
   biomedical telemetry devices.

   (b) Emissions from the device shall be confined within a 200 kHz band which
   shall lie wholly within the frequency range of 174–216 MHz.

   (c) The field strength of any emissions radiated within the specified 200
   kHz band shall not exceed 1500 microvolts/meter at 3 meters. The field
   strength of emissions radiated on any frequency outside of the specified 200
   kHz band shall not exceed 150 microvolts/meter at 3 meters. The emission
   limits in this paragraph are based on measurement instrumentation employing
   an average detector. The provisions in  Sec. 15.35 for limiting peak emissions
   apply.