FCC 15.513 Revised as of October 1, 2005
Goto Year:2004 |
2006
Sec. 15.513 Technical requirements for medical imaging systems.
(a) The UWB bandwidth of an imaging system operating under the provisions of
this section must be contained between 3100 MHz and 10,600 MHz.
(b) Operation under the provisions of this section is limited to medical
imaging systems used at the direction of, or under the supervision of, a
licensed health care practitioner. The operation of imaging systems under
this section requires coordination, as detailed in Sec. 15.525.
(c) A medical imaging system shall contain a manually operated switch that
causes the transmitter to cease operation within 10 seconds of being
released by the operator. In lieu of a switch located on the imaging system,
it is permissible to operate an imaging system by remote control provided
the imaging system ceases transmission within 10 seconds of the remote
switch being released by the operator.
(d) The radiated emissions at or below 960 MHz from a device operating under
the provisions of this section shall not exceed the emission levels in
Sec. 15.209. The radiated emissions above 960 MHz from a device operating under
the provisions of this section shall not exceed the following average limits
when measured using a resolution bandwidth of 1 MHz:
------------------------------------------------------------------------
Frequency in MHz EIRP in dBm
------------------------------------------------------------------------
960-1610.................................................. -65.3
1610-1990................................................. -53.3
011990-3100............................................... -51.3
3100-10600................................................ -41.3
Above 10600............................................... -51.3
------------------------------------------------------------------------
(e) In addition to the radiated emission limits specified in the table in
paragraph (d) of this section, UWB transmitters operating under the
provisions of this section shall not exceed the following average limits
when measured using a resolution bandwidth of no less than 1 kHz:
------------------------------------------------------------------------
Frequency in MHz EIRP in dBm
------------------------------------------------------------------------
1164-1240................................................. -75.3
1559-1610................................................. -53.3
------------------------------------------------------------------------
(f) There is a limit on the peak level of the emissions contained within a
50 MHz bandwidth centered on the frequency at which the highest radiated
emission occurs, f[M]. That limit is 0 dBm EIRP. It is acceptable to employ
a different resolution bandwidth, and a correspondingly different peak
emission limit, following the procedures described in Sec. 15.521.
[ 68 FR 19751 , Apr. 22, 2003]
CiteFind - See documents on FCC website that
cite this rule
Want to support this service?
Thanks!
Report errors in
this rule. Since these rules are converted to HTML by machine, it's possible errors have been made. Please
help us improve these rules by clicking the Report FCC Rule Errors link to report an error.