Goto Section: 95.2567 | 95.2571 | Table of Contents
FCC 95.2569
Revised as of September 1, 2021
Goto Year:2020 |
2022
§ 95.2569 MedRadio field strength measurements.
Compliance with MedRadio equivalent isotropic radiated power (M-EIRP)
limits can be determined by measuring the radiated field strength from
the transmitter type, in accordance with the rules in this section.
(a) Radiated field strength values corresponding to the M-EIRP limits
in § 95.2567 are given in the table in this paragraph, for an open area
test site, and for a test site equivalent to free space, such as a
fully anechoic test chamber. Field strength is measured at a distance
of 3 meters from the equipment under test.
M-EIRP
limit Open
area
(mV/m) Free
space
(mV/m)
1 mW 115.1 57.55
25 μW 18.2 9.1
250 nW 1.8 0.9
100 nW 1.2 0.6
(b) Compliance with the maximum transmitter power requirements in
§ 95.2567 is based on measurements using a peak detector function and
measured over an interval of time when transmission is continuous and
at its maximum power level. In lieu of using a peak detector function,
measurement procedures that have been found to be acceptable to the FCC
in accordance with § 2.947 of this chapter may be used to demonstrate
compliance.
(c) For a MedRadio transmitter intended to be implanted in a human
body, radiated emissions and M-EIRP measurements for transmissions by
stations authorized under this section may be made in accordance with
an FCC-approved human body simulator and test technique. Guidance
regarding SAR measurement techniques dielectric parameters for the
tissue-equivalent material can be found in the Office of Engineering
and Technology (OET) Laboratory Division Knowledge Database (KDB).
Goto Section: 95.2567 | 95.2571
Goto Year: 2020 |
2022
CiteFind - See documents on FCC website that
cite this rule
Want to support this service?
Thanks!
Report errors in
this rule. Since these rules are converted to HTML by machine, it's possible errors have been made. Please
help us improve these rules by clicking the Report FCC Rule Errors link to report an error.
hallikainen.com
Helping make public information public