Goto Section: 95.2563 | 95.2567 | Table of Contents
FCC 95.2565
Revised as of September 1, 2021
Goto Year:2020 |
2022
§ 95.2565 MedRadio frequency accuracy.
Each MedRadio transmitter type must be designed to maintain a frequency
stability of ±100 ppm of the operating frequency over the applicable
temperature range set forth in this section. Frequency stability
testing shall be performed over the appropriate temperature range.
(a) 25 °C to 45 °C in the case of medical implant transmitters; and
(b) 0 °C to 55 °C in the case of MedRadio programmer/control
transmitters and medical body-worn transmitters.
Goto Section: 95.2563 | 95.2567
Goto Year: 2020 |
2022
CiteFind - See documents on FCC website that
cite this rule
Want to support this service?
Thanks!
Report errors in
this rule. Since these rules are converted to HTML by machine, it's possible errors have been made. Please
help us improve these rules by clicking the Report FCC Rule Errors link to report an error.
hallikainen.com
Helping make public information public