§ 95.1209 Permissible communications.
(a) Except for the purposes of testing and for demonstrations to health
care professionals, MedRadio programmer/control transmitters may
transmit only non-voice data containing operational, diagnostic and
therapeutic information associated with a medical implant device or
medical body-worn device that has been implanted or placed on the
person by or under the direction of a duly authorized health care
professional.
(b) Except as provided in § 95.627(b) no MedRadio implant or body-worn
transmitter shall transmit except in response to a transmission from a
MedRadio programmer/control transmitter or in response to a non-radio
frequency actuation signal generated by a device external to the body
with respect to which the MedRadio implant or body-worn transmitter is
used.
(c) MedRadio programmer/control transmitters may be interconnected with
other telecommunications systems including the public switched
telephone network.
(d) For the purpose of facilitating MedRadio system operation during a
MedRadio communications session, as defined in § 95.627, MedRadio
transmitters in the 401-406 MHz band may transmit in accordance with
the provisions of § 95.627(a) for no more than 5 seconds without the
communications of data; MedRadio transmitters may transmit in
accordance with the provisions of § 95.627(b)(2) and (b)(3) for no more
than 3.6 seconds in total within a one hour time period; and MedRadio
transmitters may transmit in accordance with the provisions of
§ 95.627(b)(4) for no more than 360 milliseconds in total within a one
hour time period.
(e) MedRadio programmer/control transmitters may not be used to relay
information in the 401-406 MHz band to a receiver that is not included
with a medical implant or medical body-worn device. Wireless
retransmission of information intended to be transmitted by a MedRadio
programmer/control transmitter or information received from a medical
implant or medical body-worn transmitter shall be performed using other
radio services that operate in spectrum outside of the 401-406 MHz
band.
(f) MedRadio programmer/control transmitters and medical implant
transmitters may not be used to relay information in the 413-419 MHz,
426-432 MHz, 438-444 MHz, and 451-457 MHz bands to a receiver that is
not part of the same Medical Micropower Network. Wireless
retransmission of information to a receiver that is not part of the
same Medical Micropower Network must be performed using other radio
services that operate in spectrum outside of the 413-419 MHz, 426-432
MHz, 438-444 MHz, and 451-457 MHz bands. Not withstanding the above
restrictions, a MedRadio programmer/control transmitter of an MMN may
communicate with the MedRadio programmer/control transmitter of another
MMN to coordinate transmissions so as to avoid interference between the
two MMNs.
(g) Medical body-worn transmitters may relay only information in the
2360-2400 MHz band to a MedRadio programmer/control transmitter or
another medical body-worn transmitter device that is part of the same
Medical Body Area Network (MBAN). A MedRadio programmer/control
transmitter may not be used to relay information in the 2360-2400 MHz
band to other MedRadio programmer/controller transmitters. Wireless
retransmission of all other information from an MBAN transmitter to a
receiver that is not part of the same MBAN shall be performed using
other radio services that operate in spectrum outside of the 2360-2400
MHz band. Notwithstanding the above restriction, a MedRadio
programmer/control transmitter in the 2360-2400 MHz band may
communicate with another MedRadio programmer/control transmitter in the
2360-2400 MHz band to coordinate transmissions so as to avoid
interference between the two Medical Body Area Networks.
(h) MedRadio programmer/control transmitters operating in the 413-419
MHz, 426-432 MHz, 438-444 MHz, and 451-457 MHz bands shall not transmit
with a duty cycle greater than 3 percent.
[ 74 FR 22709 , May 14, 2009, as amended at 75 FR 52477 , Aug. 26, 2010;
77 FR 4269 , Jan. 27, 2012; 77 FR 55733 , Sept. 11, 2012; 79 FR 60100 ,
Oct. 6, 2014]
return arrow Back to Top
Goto Section: 95.1207 | 95.1211
Goto Year: 2015 |
2017
CiteFind - See documents on FCC website that
cite this rule
Want to support this service?
Thanks!
Report errors in
this rule. Since these rules are converted to HTML by machine, it's possible errors have been made. Please
help us improve these rules by clicking the Report FCC Rule Errors link to report an error.
hallikainen.com
Helping make public information public