§ 15.241 Operation in the band 174-216 MHz.
(a) Operation under the provisions of this section is restricted to
biomedical telemetry devices.
(b) Emissions from the device shall be confined within a 200 kHz band
which shall lie wholly within the frequency range of 174-216 MHz.
(c) The field strength of any emissions radiated within the specified
200 kHz band shall not exceed 1500 microvolts/meter at 3 meters. The
field strength of emissions radiated on any frequency outside of the
specified 200 kHz band shall not exceed 150 microvolts/meter at 3
meters. The emission limits in this paragraph are based on measurement
instrumentation employing an average detector. The provisions in
§ 15.35 for limiting peak emissions apply.