   (a) Operation under the provisions of this section is restricted to
   biomedical telemetry devices.

   (b) Emissions from the device shall be confined within a 200 kHz band
   which shall lie wholly within the frequency range of 174-216 MHz.

   (c) The field strength of any emissions radiated within the specified
   200 kHz band shall not exceed 1500 microvolts/meter at 3 meters. The
   field strength of emissions radiated on any frequency outside of the
   specified 200 kHz band shall not exceed 150 microvolts/meter at 3
   meters. The emission limits in this paragraph are based on measurement
   instrumentation employing an average detector. The provisions in
   § 15.35 for limiting peak emissions apply.