Goto Section: 95.1219 | 95.1301 | Table of Contents
FCC 95.1221
Revised as of October 1, 2009
Goto Year:2008 |
2010
§ 95.1221 RF exposure.
MedRadio medical implant or medical body-worn transmitters (as defined
in appendix 1 to subpart E of part 95 of this chapter) are subject to
the radiofrequency radiation exposure requirements specified in
§ § 1.1307 and 2.1093 of this chapter, as appropriate. Applications for
equipment authorization of implant devices operating under this section
must contain a finite difference time domain (FDTD) computational
modeling report showing compliance with these provisions for
fundamental emissions. The Commission retains the discretion to request
the submission of specific absorption rate measurement data.
Subpart JāMulti-Use Radio Service (MURS)